QuickSite CRT Lead Conductor Externalization: A Short Story
Sam G. Aznaurov, MD, Jeffrey N. Rottman, MD
Vanderbilt University, Nashville, Tennessee, USA
Abstract
Conductor externalization (CE) is a known phenomenon in the St. Jude Medical QuickSite and QuickFlex families of leads. This is generally considered a benign finding, not associated with electrical abnormality. We report a case of a 1056T QuickSite CRT lead CE, leading to electrical failure.
All comments will be screened and reviewed before posting. Statements, opinions, and results of studies published in AGR are those of the authors and do not reflect the policy or position of The Journal and Science International and the Editorial Board and provides no warranty as to their accuracy or reliability. Material is copyrighted and owned by Science International and cannot be used without expressed permission.